Research Instruments unveils new
version of its market leading pH meter.
NEW
FEATURES
-
Completely
new probe design
-
Improved
calibration response and repeatability
-
Improved
temperature accuracy
-
Short
term drift eliminated
-
New
firmware improves long term stability
-
Works
with all common incubators including Minc
-
All
new auto-archiving data capture software
Why
use the RI pH Meter?
The
RI pH Meter provides accurate, real-time analysis of media pH within the
incubator using its unique custom IVF pH probe. The probe fits into all
incubators and allows embryologists to adjust the CO2 concentration to obtain
the correct pH in the media. No more guesswork or assumptions when trying to
deliver the correct pH for the media being used. Measuring the pH of culture
media accurately is extremely difficult because the pH of media droplets under
oil shifts above 7.4 within just 2mins 30 seconds of removal from the incubator.
So the likelihood is, if measuring the pH of media outside of the incubator,
that you are measuring artefact pH Levels.
How
is it used?
A
media droplet is made up in the pH probe well and placed in the incubator. The
thin ribbon cable can be passed out through a sample port or through the
incubator door. The pH meter will then measure the pH in real-time displaying
the value on the digital readout.
How
does it work?
The
probe has a semiconductor sensor at the tip. The probe is placed in a holder
with a central well which allows a media/oil sample to be placed on top of it.
The probe contains a reference electrode that produces a voltage dependant on
the pH of the solution. The probe also contains a thermistor to measure the
temperature of the sample and compensate the reading.
Software
Trend Analysis
As an optional extra the meter can
have onboard memory for datalogging. Test results stored by the pH meter may be
easily downloaded to a personal computer for analysis and printing. With the new
auto archiving software, the software is simpler and easier to use.
List of worldwide distributors and further information can be found on our website
www.research-instruments.com |