Most IVF units sit on cylinder or portable vessel supply, and most inherited that arrangement rather than choosing it. It is worth reviewing, because supply mode and delivery frequency between them set a cost the unit carries every week for as long as the samples are stored.
The forecasting question matters more than the mode. A cryogenic vessel loses nitrogen continuously whether anybody opens it or not, and that standing loss typically dominates the losses caused by handling. A unit that reduces its access frequency often sees very little change in its nitrogen bill, because the work was never the expensive part.
For an embryology laboratory the practical steps are these. Establish a baseline in litres per week, in your own room, under your own workload, once the vessel has settled. Decide in advance the level at which you will act, bearing in mind that in vapour phase storage that level is higher than most people assume: samples sit above the liquid and the temperature gradient up the vessel worsens as the level falls. Then set delivery frequency so that a missed or delayed delivery still leaves headroom, with particular attention to bank holiday weekends, when levels run down unattended.
Log consumption per vessel rather than per site. A rising trend at unchanged workload is the earliest practical sign that a vacuum is degrading, and it typically appears months before frost becomes visible on the outer shell.
When reviewing the supply agreement, confirm routine and urgent lead times, minimum order and any vessel rental, who owns the vessels and who is responsible for their periodic examination, what cover exists over bank holidays, and the named escalation route when a delivery fails.